Aerosol Generator
RIGEL RI8401
HEPA filter integrity testing
- Smallest Aerosol Generator
- Use ISO 14644-3:2019 compliant aerosol type
- Remote Control to generate Aerosol
Contact us to learn more about the RIGEL 8401 Aerosol Generator
Aerosol Generator
RIGEL RI8401
The Rigel RI8401 Thermal Aerosol Generator is the perfect choice to test and validate all kinds of areas.
This is the smallest and lightest aerosol generator on the market. Plus, having a easy-drainable tank due to a valve located at the bottom makes this unit the most practical unit you can work with. The RI8401 has the ability to start the generation with a remote control (optional).
The RI8401 has a great autonomy in continuous duty: at a median concentration it can generate up to 8 hours.
No need for special kind of oil to work with this unit: it generates perfectly with the most common fluid such Emery, Ondina, PAO4 and DEHS.
This unit starts to generate a consistent amount of aerosol at very low pressures (5 PSI/0.35 bar), giving the possibility to work even in very small areas.
RIGEL RI8401 gives its best in cleanrooms , operating rooms, safety cabinets, isolators and rabs.
Aerosol Generator
RIGEL RI8401
Technical specifications
- Smallest Aerosol Generator
- Use ISO 14644-3:2019 compliant aerosol type
- Remote Control to generate Aerosol
Type of device | Aerosol generator produced by condensation of oil thermally vaporised |
Field of use | HEPA filter integrity tests (ISO 14644-3 – Section 3.2.1) |
System flow rates | 850 m3/h ÷ 119.000 m3/h (500 cfm ÷ 70.000 cfm )
|
Aerosol Concentration (at max oil output capacity) | -10μg/l @ 70,000 cfm
-100μg/l @ 7,000 cfm
|
Oil output capacity (typical)
Oil consumption
| -Max 340 mg/sec @8 bar
-Min 13 mg/sec @0.35 bar
|
Run time (Duration) | -974 min @ Min output (Min oil consumption)
-38 min @ Max output (Max oil consumption)
|
Compatible oil type (*) | PAO (**) – Durasyn 164 – Emery 3004 – Ondina L
– Smoke Oil 180 – DEHS – Smoke Oil 135 – Ondina E L
|
Thermal chamber temperature: | Programmable with thermal regulator (max 450 C)
(the temperature to be set depends on the type of oil to be used)
|
Oil tank capacity | 1 Liter |
Compressed service gas | -Type of gas: Nitrogen, CO2
-Max pressure: 8 bar
-Type of coupling: quick coupling
|
Thermal chamber temperature: | Programmable with thermal regulator
(the temperature to be set (max 380° C) depends on the type of oil to be used)
|
Pre-heating time | Min 5 min (< 10 minutes) |
Operating cycle (Duty cycle) | Continuous |
Timer module | to set specific smoke on/off times or delay the start of smoke operation |
Electrical supply | Selectable
110-120 V 50-60 Hz 10 A
220-240 V 50-60 Hz 5 A
|
Electrical protection | Internal fuse 250V 12.5 A (Appliance must be be earthed) |
Ambient conditions
during operation
| Temperature from 0°C to + 40°C
Relative humidity < 95%
|
Ambient storage conditions | Temperature from 20°C to + 55°C
Relative humidity < 95% not condensing
|
Dimensions (L x D x H) | 152 x 320 x 268 mm |
Weight | 8.9 kg (empty) |